7COMPONENT TEST(XX1, XX5 only)Test Voltage: Max. 8.6Vrms (open circuit)Test Current: Max. 11mA (shorted)Test Frequency: Line Frequency.Components: Cap

Pagina 3 - Table Of Contents:

825MHz SpecificationsCATHODE RAY TUBEVERTICAL DEFLECTIONBandwidth: DC~25 MHz(-3dB)Sensitivity: 1mV/DIV~1V/DIV(10MHz, -3dB), x5 gain selected.5mV/DIV~5

Pagina 4 - Test Instru ment Safety

9MAIN TIME BASESweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.Accuracy: ±3%±Sweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.

Pagina 5 - TEST INSTRUMENT SAFETY

COMPONENT TEST(XX1,XX5 only)Test Voltage: Max. 8.6Vrms (open circuit)Test Current: Max. 11mA (shorted)Test Frequency: Line Frequency.Components: Capac

Pagina 6 - FEATURES

40MHz SpecificationsCATHODE RAY TUBEVERTICAL DEFLECTIONBandwidth: DC~40 MHz(-3dB)Sensitivity: 1mV/DIV~1V/DIV(10MHz, -3dB), x5 gain selected.5mV/DIV~5V

Pagina 7

MAIN TIME BASESweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.Accuracy: ±3%±Sweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.A

Pagina 8 - 20MHz Specifications

COMPONENT TEST( XX1, XX5 only)Test Voltage: Max. 8.6Vrms (open circuit)Test Current: Max. 11mA (shorted)Test Frequency: Line Frequency.Components: Cap

Pagina 9

50MHz SpecificationsCATHODE RAY TUBEVERTICAL DEFLECTIONBandwidth: DC~50 MHz(-3dB)Sensitivity: 1mV/DIV~1V/DIV(10MHz, -3dB), x5 gain selected.5mV/DIV~5V

Pagina 10 - OTHER SPECIFICATIONS

MAIN TIME BASESweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.Accuracy: ±3%±Sweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.A

Pagina 11 - 25MHz Specifications

COMPONENT TEST( XX1, XX5 only)Test Voltage: Max. 8.6Vrms (open circuit)Test Current: Max. 11mA (shorted)Test Frequency: Line Frequency.Components: Cap

Pagina 13 - CH2 OUTPUT(XX0, XX5 only)

60MHz SpecificationsCATHODE RAY TUBEVERTICAL DEFLECTIONBandwidth: DC~60 MHz(-3dB)Sensitivity: 1mV/DIV~1V/DIV(10MHz, -3dB), x5 gain selected.5mV/DIV~5V

Pagina 14 - 40MHz Specifications

MAIN TIME BASESweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.Accuracy: ±3%±Sweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.A

Pagina 15

COMPONENT TEST( XX1, XX5 only)Test Voltage: Max. 8.6Vrms (open circuit)Test Current: Max. 11mA (shorted)Test Frequency: Line Frequency.Components: Cap

Pagina 16

100MHz SpecificationsCATHODE RAY TUBEVERTICAL DEFLECTIONBandwidth: DC~100 MHz(-3dB)Sensitivity: 1mV/DIV~1V/DIV(20MHz, -3dB), x5 gain selected.5mV/DIV~

Pagina 17 - 50MHz Specifications

MAIN TIME BASESweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.Accuracy: ±3%±Sweep Speed: 20nS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.Ac

Pagina 18

COMPONENT TEST( XX1, XX5 only)Test Voltage: Max. 8.6Vrms (open circuit)Test Current: Max. 11mA (shorted)Test Frequency: Line Frequency.Components: Cap

Pagina 19

1. Unpack the oscilloscope: After receipt of the instrument, immediately unpack and inspect it for any shipping damage or missing accessories. If any

Pagina 20 - 60MHz Specifications

OSCILLOSCOPE INSTRUCTION MANUAL4. Suggestions for successful instrument operation:Never place heavy objects on the instrument.Never place a hot solder

Pagina 21

2532 31 30 29 28 27 2632 25 24 23 22 21 19 1815161413101197852 4120~60MHz Standard Type Front Panel Controls and IndicatorsOSCILLOSCOPE INSTRUCTION MA

Pagina 22

26CONTROLS AND INDICATORS32 31 30 29 28 27 2632 25 24 23 22 21 19 181615141310 1197652 41 3 8 12172020~60MHz Delay Type Front Panel

Pagina 23 - 100MHz Specifications

Table Of Contents:Test Instrument SafetyGeneralNotice Before OperationOperating InstructionsApplicationMaintenanceBlock DiagramDescriptionFeatures20M

Pagina 24

2732 31 30 29 28 27 2632 25 24 23 22 21 19 18151614131011933852 41 7100MHz Standard Type Front Panel CONTROLS AND INDICATORS

Pagina 25

2832 31 30 29 28 27 2632 25 24 23 22 21 19 18161514131011933852 41 717100MHz Delay Type Front Panel CONTROLS AND INDICATORS

Pagina 26 - Notic e Before Operatio n

2937363540Standard Type Rear Panel CONTROLS AND INDICATORS

Pagina 27

30373635383940Delay Type Rear Panel CONTROLS AND INDICATORS

Pagina 28 - 15161413101197852 41

31!011. Symbol Indicators:2. Operating Controls, Indicators and Signal input connectors: (P.25~30)GENERAL FUNCTION CONTROLSON Indicator(32): Lights w

Pagina 29 - 1615141310 1197652 41 3 8 12

32CONTROLS AND INDICATORSCAL Terminal(9): Terminal provides 2V p-p, 1 KHz(nominal) square wave signal. This signal is useful for checking probe c

Pagina 30 - 151614131011933852 41 7

33CONTROLS AND INDICATORSCH 1 (X) VOLTS/DIV Control(4): Vertical attenuator for channel 1. Provides step adjustment of vertical sensitivity.

Pagina 31 - 161514131011933852 41 7

34CONTROLS AND INDICATORSPULL X5 MAG: When pulled out, increases vertical sensitivity by a factor of five. Effectively provides two extra sensitivity

Pagina 32 - Standard Type Rear Panel

MIX: The main and delayed sweep share a single trace; main sweep occupies the left portion of the display; delayed sweep occupies the right portion of

Pagina 33 - Delay Type Rear Panel

EXT: Signal from EXTERNAL TRIGGER jack becomes sweep trigger.Trigger COUPLING Switch(24): Selects trigger coupling. Four-position level switch with th

Pagina 34 - C

1Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing must often be performed where exposed

Pagina 35 - VERTICAL CONTROLS

Operating Instru ctionsOSCILLOSCOPE INSTRUCTION MANUAL371. When he oscilloscope is used to make measurements in equipment that contains high voltage,

Pagina 36 - CONTROLS AND INDICATORS

27382. Do not obstruct the ventilating holes in the case, as this will increase the scope's internal temperature.3. Excessive voltage applied to

Pagina 37 - HORIZONTAL CONTROLS

39OPERATING INSTRUCTIONSOPERATING TIPSThe following recommendations will help obtain the best performance from the oscilloscope.1. Always us

Pagina 38 - TRIGGERING CONTROLS

40OPERATING INSTRUCTIONSAll POSITION controls and INTENSITY control centered(pointers facing up).Main Time Base CONTROL TO 1 mS/DIV.Sweep Mode switch

Pagina 39 - REAR PANEL CONTROLS (P.29~30)

41OPERATING INSTRUCTIONS1. Connect probes to both the CH 1(X) and CH 2(Y) input jacks.2. Connect the ground clips of the probes to the chassi

Pagina 40 - Operating Instru ctions

42OPERATING INSTRUCTIONSTRIGGERINGThe oscilloscope provides versatility in sync triggering for ability to obtain a stable, jitter-free display in sing

Pagina 41 - OPERATING INSTRUCTIONS

43OPERATING INSTRUCTIONSTrigger SOURCE SwitchThe trigger SOURCE switch (CH 1, CH 2, etc.) selects the signal to be used as the sync trigger.1. If the

Pagina 42 - INITIAL STARTING PROCEDURE

44OPERATING INSTRUCTIONSHOLD OFF ControlControls hold-off time between sweep signals to obtain stable display when triggering a periodic sig

Pagina 43 - DUAL TRACE DISPLAY

45OPERATING INSTRUCTIONS2. Sweep MagnificationWhen a certain position of the displayed waveform is needed to be expanded time wise, a faster sweep spe

Pagina 44

46OPERATING INSTRUCTIONSThe portion of A is main sweep which is set by the MAIN TIMEBASE. The portion of B is delay sweep witch is set by the

Pagina 45

7. Some equipment with a two-wire ac power cord, including some with polarized power plugs, is the "hot chassis" type. This includes most

Pagina 46 - SLOP Switch

47WAVEFORM CHARTREDB CEBLKREDB CEBLK1000 ΩREDB CEBLK0 (short)Ω 500 resistorΩ 10K resistorΩ1 F ceramic capacitorμ 1 F electrolytic μ capacitor 10

Pagina 47 - TIME BASE Control

ApplicationOSCILLOSCOPE INSTRUCTION MANUAL48DC VOLTAGE MEASUREMENTS(Refer to Fig.7)The following technique may be used to measure the instantaneous de

Pagina 48 - Start point C mixed point

49APPLICATIONMEASUREMENT OF VOLTAGE BETWEEN TWO POINT ON A WAVEFORM(Refer to Fig.8)This procedure may be used to measure peak-to peak voltages, o

Pagina 49 - COMPONENT TEST

50APPLICATIONFir the example shown in Fig.8, the two points are separated by 4.4 divisions vertically. If the VOLTS/DIV setting is 20 mV

Pagina 50 - WAVEFORM CHART

51APPLICATIONFor the example shown in Fig.9, the horizontal distance between the two points is 5.4 divisions. If the TIME/DIV is 0.2 ms and X10 magnif

Pagina 51 - Application

52APPLICATIONMethod No.2(Refer to Fig.11)While the previously described method relies upon direct period measurement of one cycle, the frequency may a

Pagina 52 - ON A WAVEFORM

53APPLICATIONPULSE WIDTH MEASUREMENTS(Refer to Fig.12)1. Apply the pulse signal to the input jack and set the CH 1/CH 2 switch to the channel

Pagina 53 - TIME MEASUREMENTS

54APPLICATIONPULSE RISE TIME AND FALL TIME MEASUREMENTS(Refer to Fig.13)For rise time and fall time measurements, the 10% and 90% amplitude point are

Pagina 54 - FREQUENCY MEASUREMENTS

55APPLICATIONTIME DIFFERENCE MEASUREMENT(Refer to Fig.14)This procedure is useful in measurement of time difference between signals that are synchroni

Pagina 55 - Method No.2

56APPLICATIONPHASE DIFFERENCE MEASUREMENTSMethod No.1(Refer to Fig.15)This procedure is useful in measuring the phase difference of signals of the sam

Pagina 56 - PULSE WIDTH MEASUREMENTS

3GeneralOSCILLOSCOPE INSTRUCTION MANUALDESCRIPTIONThe model PS-XX0/XX1/XX5 oscilloscope is a dual-channel oscilloscope with frequency bandwidt

Pagina 57 - Rise time

57APPLICATIONMethod No.2(Refer to Fig.16)The above procedure allows 45 per division, which may not give the desired accuracy for small ph

Pagina 58 - Time difference

58APPLICATIONX-Y MODE APPLICATIONS(Refer to Fig.17)A dual-trace method of phase measurement was previously described. A second method of phases mea

Pagina 59 - PHASE DIFFERENCE MEASUREMENTS

59APPLICATIONFREQUENCY RESPONSE MEASUREMENTS(Refer to Fig.19)A sweep generator and the X-Y mode of this oscilloscope may be used to measure the audio

Pagina 60 - One cycle adjusted to 8 div

MaintenanceOSCILLOSCOPE INSTRUCTION MANUAL60A following instruction are for use by qualified service personnel only. To avoid electrical

Pagina 61 - X-Y MODE APPLICATIONS

61MAINTENANCE2. Set the probe to X10 (compensation adjustment is not possible in the X1 position).3. Touch tip of probe to CAL terminal.4. Adjust osc

Pagina 62 - APPLICATION

Block DiagramOSCILLOSCOPE INSTRUCTION MANUAL62CH A ATTCH B ATTCH APRE AMPTRIG, X-AIXS PICK OFF AMPCH BPRE AMPCOMP TESTVERT MODE SWITCHBEAM FINDER SW

Pagina 65 - Block Diagram

TINSE0049S4 Ver.01

Pagina 66

4GENERAL(9) Z Modulation (PS-XX5 only)Input terminal is used for external intensity modulation signal.(10) Component Test (PS-XX1/XX5)Capacitor, indic

Pagina 67

520MHz SpecificationsCATHODE RAY TUBEVERTICAL DEFLECTIONBandwidth: DC~20 MHz(-3dB)Sensitivity: 1mV/DIV~1V/DIV(5MHz, -3dB), x5 gain selected.5mV/DIV~5V

Pagina 68 - TINSE0049S4 Ver.01

6MAIN TIME BASESweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.Accuracy: ±3%±Sweep Speed: 0.1μS/DIV~2.0S/DIV in 1-2-5 sequence, 23 steps.